aep MP1
Panel digital indicator designed to be used in the modern static and dynamic measure systems in industrial activity, it can sample the signal up to a frequency of 300Hz.
Suitable for processing analogue signals coming from strain gauge based load cells and from pressure transmitters with amplified output, it enables to perform measurements with a long-term accuracy of 0.05% of full scale, with a resolution of ±9.999 divisions 2mV/V.
Accuracy 0.05%
Suitable for processing analogue signals coming from strain gauge based load cells and from pressure transmitters with amplified output, it enables to perform measurements with a long-term accuracy of 0.05% of full scale, with a resolution of ±9.999 divisions 2mV/V.
Accuracy 0.05%
Product Description
DIN 43750 case
2 SET POINTS: with relay output, activation and deactivation thresholds (hysteresis) programmable.Digital FILTER: it allows to attenuate the measurement oscillations in dynamic applications.RESOLUTION: it allows to increase the measurement by definite steps.(1,2,5 ….).PEAK: programmable, it allows to detect measurement peaks in positive or negative field.ZERO key BLOCK: it allows to block the Zero key function by password.REMOTE INPUT digital: programmable by menu to perform a function to be chosen among Peak, Zero and Hold
2 SET POINTS: with relay output, activation and deactivation thresholds (hysteresis) programmable.Digital FILTER: it allows to attenuate the measurement oscillations in dynamic applications.RESOLUTION: it allows to increase the measurement by definite steps.(1,2,5 ….).PEAK: programmable, it allows to detect measurement peaks in positive or negative field.ZERO key BLOCK: it allows to block the Zero key function by password.REMOTE INPUT digital: programmable by menu to perform a function to be chosen among Peak, Zero and Hold