KOEHLER K25900 Constant Temperature Pressure Hydrometer Water Bath |
Test Method Density and relative density measurements of light hydrocarbons, including LPG, are used for transportation, storage and regulatory purposes. The measurement is made by floating a thermohydrometer in a sample that has been introduced into a pressure cylinder. Features and Benefits • Conforms to ASTM D1657 and related specifications • Mechanically refrigerated for convenient sub-ambient temperature operation |
K25900 Constant Temperature Pressure Hydrometer Water Bath Immerses two Pressure Hydrometer Cylinders at 60°F (15°C) for density and relative density determinations of LPG and other light hydrocarbons. Mechanically refrigerated cooling system maintains sub-ambient temperature. Thermistor activated solid state temperature controller and 750W copper immersion heater maintain bath temperature with ±0.5°F (±0.2°C) stability. A 1/20 hp ball bearing stirrer circulates the bath medium to assure temperature uniformity. Stainless steel tank is fiberglass insulated. Equipped with overflow standpipe/drain. Steel exterior has a durable polyurethane enamel finish. Specifications Conforms to the specifications of: ASTM D1657; IP 235; ISO 3993 Controller Sensitivity: ±0.5°F (±0.2°C) Capacity: Two (2) K26150 cylinders Electrical Requirements 115V 60Hz, Single Phase, 12.5A 220-240V 50 or 60Hz, Single Phase, 6.4A Dimensions lxwxh,in.(cm) Bath interior: 12x18x22(30x46x56) Overall: 18x20x49 (46x51x124) Net Weight: 158 lbs (71.7kg) Shipping Information Shipping Weight: 186 lbs (84.4kg) Dimensions: 15.4 Cu. ft. |
Ordering Information | |||||||||||||
|