KOEHLER K33850 Existent Gum Steam Generator |
Test Method Gum formed during fuel storage can deposit on induction system surfaces, intake valves, stems and guides. To test for gum content, a 50mL sample is evaporated in an aluminum block bath for a specified period under controlled conditions of temperature and flow of air (aviation and motor gasolines) or steam (aircraft turbine fuel). Features and Benefits • For steam-jet method testing of aircraft turbine fuels • Meets output requirements of three-unit and six-unit evaporation baths • Electrically heated for clean, efficient operation and ease of installation • Meets applicable ASME, NEC standards; UL listed, CSA approved |
K33850 Existent Gum Steam Generator Electrically heated boiler provides instantaneous and reserve steam capacity for steam-jet evaporation tests. Easy to install and operate; electrical heating eliminates the need for on-site fuel combustion. Requires only a water feed source and electrical hook-up. Ruggedly constructed, with long life industrial grade incoloy heating element. Includes a full range of safety features: automatic water level control and low water cut-off; steam safety valve; high-limit pressure cut-out with manual reset; steam pressure gauge. Specifications Output: 54.1 lbs steam/hr. 212°F Bhp Rating: 1.83 kW Rating: 18 Dimensions lxwxh,in.(cm) 20x28x36 (51x71x91) Net Weight: 185 lbs (83kg) Shipping Information Shipping Weight: 200 lbs (91kg) Dimensions: 18 Cu. ft. |
Ordering Information | |||||||||||||||||||||||||
|
Accessories | |||||||||||||
|