HANNA FC100B Foodcare pH Electrode

HANNA FC100B Foodcare pH Electrode for the Food Industry









The FC100B is a PVDF body, refillable, double junction pH electrode with a BNC connector. This single ceramic junction electrode has a pH indicating probe made of general purpose glass and a food grade plastic body. This design consideration is ideal for pH measurements in the food industry including sauces, juices, dairy products and other liquid or slurry forms of food. The recommended operating temperature range is from 0 to 80°C.
pvdf body

Durable PVDF body

single ceramic double junction

Double junction design

refillable electrode

Refillable electrode

Details

Hanna Instruments offers a wide variety of pH electrodes that are designed for many different applications. The type of glass used for sensing pH, bulb shape, body material, type of junction, type of reference and electrolyte used are just some of the design considerations.
The FC100B uses general purpose (GP) glass, spherical bulb, food grade PVDF body, single ceramic frit, double junction and is refillable with 3.5M KCl.
high temperature glass

General Purpose Glass Formulation

General purpose glass as the name implies is a standard glass formulation that is used for general use.  A pH electrode with general purpose (GP) glass will have a resistance of 100 megaohms at 25°C and is suited for measuring pH of samples that are at ambient temperatures. The FC100B is suitable to use with samples that measure from 0 to 80°C.   
high temperature glass


Spherical Bulb

The spherical bulb is for general purpose use within the food industry. Measurements can be made in liquids or slurries that allow easy measurement of solid and semi-solid food products. Other tip shapes include conic for penetration and flat tip for surface measurements.
pvdf body

 

PVDF Body

Polyvinylidene fluoride (PVDF) is a food grade plastic that is resistant to most chemicals and solvents, including sodium hypochlorite. It has high abrasion resistance, mechanical strength and resistance to ultraviolet and nuclear radiation. PVDF is also resistant to fungal growth.
double junction reference
   


Single Ceramic Double Junction Reference

The FC100B is a double junction design. pH electrodes are available as single junction or double junction. See below for a complete description of the differences. The junction also known as a salt bridge is necessary component of the electrical circuit. The movement of ions must flow through the junction for a steady reading. The outer reference has a single ceramic frit. The ceramic is a porous material that is easily fused with the glass body and has a similar expansion coefficient. A single ceramic junction has a flow rate of 15-20 μL/hour. Other junction types are available with higher flow rates and made with different materials.
refillable  

Refillable

The FC100B is a refillable probe. Since it is a double junction pH electrode the fill solution is the HI7082 3.5M KCl. This solution does not contain any silver as with single junction electrode. The absence of silver will prevent any silver precipitate from forming at the junction surface and clogging it. Clogging of the junction will result in drifty and erratic readings.
bnc connector

BNC Connector

The FC100B uses a BNC connector. This type of connector is universal in that it can be used on any pH meter that has the female BNC probe input. Other type of connectors include DIN, screw type, T-type, and 3.5mm to name a few. These types of connectors tend to be proprietary for a particular type of meter and are not interchangeable.


Single Junction Versus Double Junction pH Electrodes

junction types

Conventional electrodes are normally single junction. As depicted by the figure above, these electrodes have only a single junction between the internal refernce wire and the external solution. Under adverse conditions, such as high pressure, high temperature, highly acidic or alkaline solutions, the positive flow of the electrolyte through the junction is often reversed resulting in the ingress of sample solution into the reference compartment. If this is left unchecked, the reference electrode can become contaminated, leading to complete electrode failure. Another potential problem with single junction electrodes is the clogging of the junction due to silver chloride (AgCl) precipitation. Silver can be easily precipitate in samples that contain Tris buffer or heavy metals. When the electrolyte solution makes contact with the sample, some AgCl will precipitate on the external face of the junction. The result is drifty readings obtained from the sensor.
Hanna’s double junction system, as the name implies, has two junctions, only one of which is in contact with the sample as shown in the figure. Under adverse conditions, the same tendency of sample ingress is evident. However, as the reference electrode system is separated physically from the intermediate electrolyte area, the contamination of the electrode is minimized. The likelihood of clogging of the junction is also reduced with a double junction electrode since the outer reference cell uses a fill solution that is “silver-free”. Since there is no silver present, there is no precipitate forming to clog the junction. 

Specifications

Body MaterialPVDF
Referencedouble, Ag/AgCl
Junction / Flow Rateceramic, single / 15-20 μL/h
Electrolyte3.5M KCl
RangepH: 0 to 13
Max Pressure0.1 bar
Tip Shapespheric (dia: 7.5 mm)
Diameter12 mm
Body Length / Overall Length120 mm / 160 mm
Recommended Operating Temperature0 to 80°C (32 to 176°F)
Temperature Sensorno
Matching Pinno
Amplifierno
Cablecoaxial; 1 m (3.3’)
ConnectionBNC
Applicationsdairy, juice, liquids and slurries measurement in food industry, sauces