Jumat, 04 Maret 2016

HANNA HI1093B pH Electrode with Micro Bulb and Extended Length

HANNA HI1093B pH Electrode with Micro Bulb and Extended Length



The HI1093B is a glass body, single reference junction pH electrode with a BNC connector designed to be specifically used to measure pH in NMR tubes. This pH electrode has a 3mm (0.11”) diameter and is 130mm (5.11”) long. The electrode has a unique open junction design in which there is a viscolene (hard gel) electrolyte layer that is free of silver chloride (AgCl) between the sample to be measured and the internal reference cell. The open junction design resists clogging and the probe’s spherical micro bulb tip made from general purpose glass.

3mm (0.11”) Micro bulb tip

130mm (5.11”) probe length

Open junction design

Details

Hanna Instruments offers a wide variety of pH electrodes that are designed for many different applications. The type of glass used for sensing pH, bulb shape, body material, type of junction, type of reference and electrolyte used are just some of the design considerations.
The HI1093B is a glass body probe that uses general purpose (GP) glass and open junction design with viscolene gel electrolyte. The diameter of the probe is 3mm (0.11”) diameter and the length is 130mm (5.11”), making it ideal for NMR tubes.

Glass

General Purpose Glass Formulation

General purpose glass as the name implies is a standard glass formulation that is used for general use.  A pH electrode with general purpose (GP) glass will have a resistance of 100 megaohms at 25°C and is suited for measuring pH of samples that are at ambient temperatures. The HI1093B is suitable to use with samples that measure from 0 to 50°C
Micro Bulb

Micro Bulb Tip

The 3 mm micro bulb tip design allows for measurement in samples less than 100 µL. In the field of biotechnology, samples may need to be measured in test tubes, vials, and NMR tubes, which the small bulb of the HI1093B easily allows for.
Glass Body 

Extended Glass Body

The glass body is ideal for laboratory use. The glass is resistant to many harsh chemicals and is easily cleaned. The glass body also allow for a fast transfer of heat to the internal reference electrolyte. The mV generated by the reference cell is temperature dependent. The faster the equilibrium the steadier the reference potential. The HI1093B extended body length allows for measurement in NMR tubes.

Open Junction

Open Junction Reference

Suspended solids and proteins found in food products will clog a conventional ceramic reference junction. This clogging will impede the measurement circuit between the indicating electrode and the internal reference resulting in slower response time, erratic readings, and frequent electrode replacement. The open junction design consists of a solid gel interface (viscolene) between the sample and internal Ag/AgCl reference. This interface not only prevents silver from entering the sample, but also makes it impermeable to clogging, resulting in a fast response and stable readings.

din connector

 BNC Connector

The HI1093B uses a BNC connector. This type of connector is universal in that it can be used on any pH meter that has the female BNC probe input.  Other type of connectors include DIN, screw type, T-type, and 3.5mm to name a few. These types of connectors tend to be proprietary for a particular type of meter and are not interchangeable.

Specifications

Body Materialglass
Referencesingle, Ag/AgCl
Junction / Flow Rateopen
Electrolyteviscolene
RangepH: 0 to 13
Max Pressure0.1 bar
Tip Shapespheric (dia: 3 mm)
Diameter3 mm
Body Length / Overall Length130 mm / 173 mm
Recommended Operating Temperature0 to 50°C (32 to 122°F) - GP
Temperature Sensorno
Matching Pinno
Amplifierno
Digitalno
Cablecoaxial; 1 m (3.3’)
ConnectionBNC
ApplicationsNMR tubes