Jumat, 04 Maret 2016

HANNA HI1332B Combination pH Electrode

 HANNA HI1332B Combination pH Electrode for Chemicals, Quality Control and Field





The HI1332B is a PEI body, double junction pH electrode with a BNC connector. This electrode has a single ceramic junction in the outer reference cell and the pH sensing portion is made with high temperature glass, protected by an extension of the durable PEI body. This design consideration is ideal for field applications as well as general purpose use.

Durable PEI body

Double junction design

Refillable electrode

Details

Hanna Instruments offers a wide variety of pH electrodes that are designed for many different applications. The type of glass used for sensing pH, bulb shape, body material, type of junction, type of reference and electrolyte used are just some of the design considerations.
The HI1332B uses general purpose (GP) glass, spherical bulb, polyetherimide (PEI) body, single ceramic frit, and double junction.

Glass

General Purpose Glass Formulation

The measurement of pH at very high temperatures is detrimental to the sensitive glass bulb and will shorten the life of it. A pH electrode with general purpose (GP) glass will have a resistance of 100 megaohms at 25°C while the resistance of LT glass is around 50 megaohms at 25°C. As the temperature of the glass decreases in the sample, the resistance of the LT glass will approach that of GP glass. If using GP glass, the resistance would increase above the optimum range, resulting in increased impedance and ultimately affecting the measurement. The HI1332B is suitable to use with samples that measure from 0 to 70°C.

Spherical Bulb

Spherical Bulb

The spherical bulb is for general purpose use. Other tip shapes include conic for penetration and flat tip for surface measurements.
PEI Body 

PEI Body

The body of the HI1332B is composed of polyetherimide (PEI) resin. The PEI body is suitable for a wide range of applications and excels in field measurements due to its durability. The shielding around the spherical glass tip minimizes breakage due to accidental bumping or dropping of the electrode. The PEI plastic is a high quality plastic that is chemically resistant to many aggressive chemicals.

Double Junction

Double Junction Reference

A double junction electrode has an internal compartment surrounding the reference wire. Silver ions are present in the electrolyte of the internal compartment, which houses the Ag/AgCl reference wire; the electrolyte outside this compartment is silver free. The double junction design means that virtually no silver from the electrode enters the sample. This design allows measurement in applications where silver ions in the sample are undesirable or silver precipitates on the junction are likely to form.

din connector

 BNC Connector

The HI1332B uses a BNC connector. This type of connector is universal in that it can be used on any pH meter that has the female BNC probe input. Other type of connectors include DIN, screw type, T-type, and 3.5mm to name a few. These types of connectors tend to be proprietary for a particular type of meter and are not interchangeable.

single vs double junction
Conventional electrodes are normally single junction. As depicted by the figure above, these electrodes have only a single junction between the internal refernce wire and the external solution. Under adverse conditions, such as high pressure, high temperature, highly acidic or alkaline solutions, the positive flow of the electrolyte through the junction is often reversed resulting in the ingress of sample solution into the reference compartment. If this is left unchecked, the reference electrode can become contaminated, leading to complete electrode failure. Another potential problem with single junction electrodes is the clogging of the junction due to silver chloride (AgCl) precipitation. Silver can be easily precipitate in samples that contain Tris buffer or heavy metals. When the electrolyte solution makes contact with the sample, some AgCl will precipitate on the external face of the junction. The result is drifty readings obtained from the sensor.
Hanna’s double junction system, as the name implies, has two junctions, only one of which is in contact with the sample as shown in the figure. Under adverse conditions, the same tendency of sample ingress is evident. However, as the reference electrode system is separated physically from the intermediate electrolyte area, the contamination of the electrode is minimized. The likelihood of clogging of the junction is also reduced with a double junction electrode since the outer reference cell uses a fill solution that is “silver-free”. Since there is no silver present, there is no precipitate that can form to clog the junction.

Specifications

Body MaterialPEI
Referencedouble, Ag/AgCl
Junction / Flow Rateceramic, single / 15-20 μL/h
Electrolyte3.5M KCl
RangepH: 0 to 13
Max Pressure0.1 bar
Tip Shapespheric (dia: 9.5 mm)
Diameter12 mm
Body Length / Overall Length120 mm / 175.5 mm
Recommended Operating Temperature0 to 70°C (32 to 158°F) - GP
Temperature Sensorno
Matching Pinno
Amplifierno
Digitalno
Cablecoaxial; 1 m (3.3’)
ConnectionBNC
Applicationschemicals, field applications, quality control

Tidak ada komentar: