KOEHLER K10200 Automatic Aniline Point Apparatus

KOEHLER K10200 Automatic Aniline Point Apparatus



Test Method
Aniline point is used to characterize pure hydrocarbons and to indicate the aromatic content of hydrocarbon mixtures. Equal volumes of aniline and sample or sample plus n-heptane are stirred together while being heated at a controlled rate. After the two phases become miscible, the mixture is cooled at a controlled rate and the temperature at which the two phases separate is the aniline point or mixed aniline point of the sample.

Features and Benefits
• Conforms to ASTM D611 and related specifications

• For samples ranging from clear to very dark

• Temperature range 0° to 150°C (32 to 302°F)

• Digital temperature display


Click on the image to view a larger photo of the instrument.

K10200 Automatic Aniline Point Apparatus
Performs aniline point and mixed aniline point determinations automatically by means of a modified thin film technique (ASTM D611 Method E). The sample-aniline mixture is directly heated by a platinum immersion heater and the aniline point is detected photoelectrically. Temperature is displayed on a large LED indicator. Built-in pressure regulator and solenoid valve permit the use of cooling air for quicker cooling cycles or to determine subambient aniline point temperatures. Aniline points as low as 0°C (32°F) can be determined with the use of refrigerated cooling air. Equipped with variable controls for heater, light source and stirrer speed. Cabinet exterior surfaces have a chemical resistant polyurethane enamel finish.


Specifications
Conforms to the specifications of:
ASTM D611; IP 2; ISO 2977; DIN 51775; FTM 791-3601; NF M 07-021

Testing Range: 0 to 150°C (32 to 302°F)

Temperature Display: 0-999.9°C

Electrical Requirements
115V, 50/60Hz, Single Phase, 0.4A
220-240V, 50/60Hz, Single Phase, 0.2A

Included Accessories
Standard Pyrex™ Test Cell with drain

Dimensions lxwxh,in.(cm)
14.5x8.5x20.5 (37x22x53)
Net Weight: 32.5 lbs (14.7kg)

Shipping Information
Shipping Weight: 46 lbs (21kg)
Dimensions: 8.2 Cu. ft.

Ordering Information
Catalog
Number
Product Description
Electrical Requirements
K10200Automatic Aniline Point Apparatus115V 50/60Hz
K10290Automatic Aniline Point Apparatus220-240V 50/60Hz


Accessories

Catalog
Number
Product Description
Temperature Range
250-000-33F
ASTM 33F Thermometer
-36.5 to +107.5°F
250-000-33C
ASTM 33C Thermometer
-38 to +42°C
250-000-34F
ASTM 34F Thermometer
77 to 221°F
250-000-34CASTM 34C Thermometer
25 to 105°C
250-000-35FASTM 35F Thermometer
194 to 338°F
250-000-35CASTM 35C Thermometer
90 to 170°C
K10210Pyrex Test Cell with drain
K10220
Heating-Cooling Tube with platinum element

Tidak ada komentar: